MFFT 90 Minimum Film Forming Temperature Instrument

Industries & Applications

MFFT 90 Minimum Film Forming Temperature Instrument

Try before you buy

Find out more

Extended warranty

Find out more

Industries & Applications

Description

 

Paints & coatings

The minimum film forming temperature of a paint or coating is the lowest temperature at which it will uniformly coalesce when laid on a substrate as a thin film. The standard test for determining this temperature involves using an MFFT, as specified by such standards as ASTM D 2354 and ISO 2115. The design of the Rhopoint instrument is based directly on these standards.

Adhesives

The MFFT of one-pack adhesives can be found in a similar way to paints and coatings. With two-pack adhesives, a strip of aluminium foil can be laid on the platen, to ease cleaning, before applying the adhesive. Two-pack adhesives do not show an MFFT, but the minimum cure temperature can be found by scraping the sample with a spatula.

Running

The desired temperature program is selected and the instrument allowed to reach thermal equilibrium. Tracks of wet test material are applied using a cube applicator, or spreader. Once the material has dried or cured the result is visually apparent.

Standard features

Standard model (MFFT-90) has ten temperature programs:

 0123456789
Cold end ˚C
(˚F)
-10
(14)
-5
(23)
0
(32)
5
(41)
15
(59)
23
(73)
33
(91)
43
(110)
53
(127)
63
(145)
Warm end ˚C
(˚F)
8
(45)
13
(55)
18
(64)
23
(73)
33
(91)
50
(122)
60
(140)
70
(158)
80
(176)
90
(194)

Find out more about MFFT in our article ‘What is MFFT?’

The MFFT Method

A dull, nickel-plated copper platen is cooled at one end electronically and warmed at the other end, while air or nitrogen is caused to flow over the surface as a uniform blanket. The air flow is from cold end to hot end.

To achieve the required uniformity, the air or gas is delivered via a very carefully designed sintered metal distribution block. The design is such that freezing does not take place at the inlet. For use with air, a drying system is incorporated into the housing together with a flow controller. The air dryer contains indicator crystals which are clearly visible in their transparent container. The complete air conditioning system is accessible at the side of the instrument.

Water at normal mains pressure removes the excess heat from the coolers. In environments where the ambient water temperature is high, Rhopoint offer a chiller / recirculating unit to reduce this temperature. Quick release couplings are provided. Water is normally drawn from a laboratory tap and the outlet is run to drain by gravity. Alarms, both audible and visible, protect the cooling water supply.

Temperature sensors are mounted at intervals under the surface of the platen, these are used to control the temperature of the platen in accordance with the chosen program. They are also used to indicate platen temperatures down the length of the bar. A hinged viewing cover over the platen provides thermal insulation while allowing visual inspection of the determination as it progresses.

The use of this instrument may involve hazardous materials and operational procedures. It is the responsibility of the user of this instrument to establish appropriate safety and health practices and determine the applicability of regulatory limitations prior to use.

How does the MFFT work?

1. Select program

Touch-screen intuitive operation with in-built operating guide on-screen

2. Fill application cube with product

3. Move across platen and close perspex lid

4. Coalescence will show in about 90 minutes

5. Move cursor

6. Read the exact MFFT point from the display

Video

Specifications

Instrument Specifications
Conforms toASTM D 2354, ISO 2115
Platen Dimensions483mm × 235mm (L × W)
Air4 l min-1 @ 100 psig
WaterMains pressure; gravity drain
inlet temperature <15°C
Mains110–120V AC or
220–240V AC
Order CodeMFFT, -10˚C to 90˚C 230V - A1212-002
MFFT, -10˚C to 90˚C 115V - A1212-002/115
Dimensions350mm × 550mm × 610mm (H × W × D)
Weight38kg
Packed Weight53.7kg
Packed Dimensions590mm × 800mm × 690mm (H × W × D)
Commodity Code9027 8017
Accessories
Included• Water connectors
• Air connector
• Spare silica gel
• Manual
• 75μm (0.003”) cube applicator
Optional• Chiller - Required when the temperature of the water supply to the instrument is >20ºC higher than the target temperature at the cold end of the selected range.
e.g. Range 0, -10ºC requires a water supply of no higher than 10ºC. If this cannot be supplied then the chiller is required
• Label printer
Extras• Free extended warranty service
• Fast and economic service via our global network of accredited service centres, please visit www.rhopointinstruments.com/distributors/category/service-centre/ for detailed information

What’s in the box

Included:

  • Water connectors
  • Air connector
  • Spare silica gel
  • Manual
  • 75μm (0.003”) cube applicator

Optional extras

Chiller

Required when the temperature of the water supply to the instrument is >20ºC higher than the target temperature at the cold end of the selected range. e.g. Range 0, -10ºC requires a water supply of no higher than 10ºC. If this cannot be supplied then the chiller is required. View chiller

Cube applicators

Stainless Steel 1” cube applicator, 75 & 150 micron cuts
Stainless Steel 1” cube applicator, 50 & 100 micron cuts

Please ask for these extras in your quote request

Extended 2 Year Warranty
Customer Support
Servicing

Standards & Certifications

ASTM 2354

Standard Test Method for Minimum Film Formation Temperature (MFFT) of Emulsion Vehicles

ISO 2115

Plastics — Polymer dispersions — Determination of white point temperature and minimum film-forming temperature

Applications

Paints and Coatings

The minimum film forming temperature (MFFT) of a paint or coating is the lowest temperature at which it will uniformly coalesce when laid on a substrate as a thin film. The standard test for determining this temperature involves using a MFFT-BAR, as specified by such standards as ASTM D 2354 and ISO 2115. The design of the Rhopoint instrument is based directly on these standards.

Adhesives

The MFFT of one-pack adhesives can be found in a similar way to paints and coatings. With two-pack adhesives, a strip of aluminium foil can be laid on the platen, to ease cleaning, before applying the adhesive. Two-pack adhesives do not show an MFFT, but the minimum cure temperature can be found by scraping the sample with a spatula

Related News Articles

Answers to common questions about our instruments and measurements

Examples of low gloss finishes include:

Selecting the correct angle for the application will optimise measurement accuracy. Three types of instruments are available on the market: 60° single angle instruments, a

What is RSpec?

The diagram above shows the same coating applied to two panels, the first has orange peel/waviness present due to the incorrect application setting on the

Email me a price and product summary

Please note, this is NOT a quotation, just a confirmation of the price when purchased in the UK exclusive of VAT/duties

Want an email with price and product summary?

Please note, this is NOT a quotation, just a confirmation of the price when purchased in the UK exclusive of VAT/duties

MFFT 90 Minimum Film Forming Temperature Instrument