Calibrating Aesthetix in AE

Once a module is chosen, it may be required to calibrate the sensor.

Calibration Standards

The following standards must be used dependent on the module.

Module Standard
Gloss Module B8000-011 Gloss Module Standard
Texture Module B8000-012 Texture Module Standard
Coatings Physical Test B8000-011 Gloss Module Standard

Modules which do not require calibration include:

  • Sparkle Module

Click on the calibration icon to begin calibration and follow the on-screen instructions.
Calibrating an Instrument in AE

Calibration interval: The instrument should be checked by measuring the calibration tile dailyweekly and comparing read values with the certified values. If values are out of tolerance, recalibrate the sensor.
Calibration artifacts must be clean with no contamination, visible damage, or fingerprints.