Calibrate the TAMS
Calibrate the TAMS
Before taking measurements, calibrate TAMS using the supplied plate so focus and reference values are correctly calibrated.
Calibration plate
The plate includes three tiles:
- Plastic‑ref – for surface focus.
- Silver‑ref – for screen focus and reference calibration.
- Check tile‑ref – for later verification only (not used during calibration).
Start calibration
- On the instrument, select Menu → Calibration → Start calibration process.
- Confirm with YES and press OK.
Step 1 – Plastic‑ref
- Place TAMS on the Plastic‑ref tile (top tile).
- Ensure all four feet are flat on the tile.
- Select Continue and press OK.ly calibrated
TAMS now sets the surface autofocus.
Step 2 – Silver‑ref
- Move TAMS to the Silver‑ref tile (mirror‑like middle tile).
- Ensure all four feet are flat on the tile.
- Select Continue and press OK.
TAMS now sets the screen autofocus and reference calibration, then returns to normal operation.
Step 2 – Gloss
- Move TAMS to the Gloss‑ref tile.
- Ensure all four feet are flat on the tile.
- Select Continue and press OK.
TAMS now calibrates the gloss measurement.