QUICK START OPERATING GUIDE

Rhopoint IQ Flex 20 & Flex 20-S

Instrument Controls

fig.1 - Instrument controls

A detailed product manual can be found on the USB supplied with the instrument.
Or it can be downloaded from: www.rhopointinstruments.com/help-services/resources/

Basic Operations

Navigation

  • Available navigation options are displayed as prompts on the top left of every screen
  • Use the indicated button to access required menus and options
  • Press left button twice to return home from any menu.

Measurement

  • Place instrument measuring head on the surface to be measured. Ensure that the measurement spot is completely covered by the surface to be measured. For more repeatable testing, use the supplied positioning template to ensure that measurements are made in the same position across multiple samples.
  • Press a measurement button (as shown in fig.1) to take a measurement.

Statistics

  • Press right button to view statistics
  • Press right button again to view graphs
  • Advanced options are available in: MENU > SETUP > BATCH MENU

Calibration

Rhopoint IQ Flex 20 calibration

  1. Ensure tile is clean and undamaged
  2. Optics should also be checked for cleanliness
  3. Place instrument measuring head on the tile holder. Ensure the correct adaptor is attached. Please note, the instrument must be recalibrated each time the adaptor is changed
  4. Ensure there is no movement of the remote head on the calibration tile
  5. Navigate to: MENU > CALIBRATE INSTRUMENT
  6. Press a measurement button (as shown in fig.1) to calibrate and select ‘confirm’ once complete

RECALIBRATION
Rhopoint recommend that the Rhopoint IQ Flex 20 should be regularly serviced and calibrated. This will ensure reliable, accurate and consistent measurements are given. For more information or to book a service / calibration please visit: www.rhopointinstruments.com/help-services/calibration-servicing/

Advanced Operations

User Defined Batches

  • Connect your instrument to a PC via a USB cable
  • Open the ‘IQ-METER’ drive in your file explorer
  • Open the ‘CONFIG’ folder
  • Edit ‘batches.txt
  • End by pressing RETURN
  • Save file

Data Transfer

  • Connect instrument to PC via USB cable
  • Open ‘IQ-METER’ USB drive
  • Open ‘DATA’ folder
  • Each stored batch has an individual folder
  • Batch Statistics are stored as ‘Stats.csv
  • Individual readings are stored separately in .csv format
  • Copy, view and delete data as required.

Automatic Measurement

  • Navigate to MENU > SETUP > Measurement Menu > Auto Measure
  • Select measurement interval of 2/5/10/30/60 seconds
  • Press to go back until the main measurement screen is displayed
  • Press to start and stop automatic measurements

Screen Navigation

qsg-rhopoint-iq-instrument-navigation-screens-english
NB. Not all screens are available in portrait orientation, refer to manual for further information
qsg-rhopoint-iq-instrument-settings-screens-english
NB. Not all screens are available in portrait orientation, refer to manual for further information

References

GLOSS
Gloss is a measure of how much light is reflected from a sample surface at a defined angle.

AngleApplicationRange
20°High gloss finishes0 - 2000 GU
60°All gloss finishes0 - 1000 GU
85°Low gloss finishes0 - 199 GU

HAZE
Reflection haze is scattered light caused by micro texture and is measured adjacent to the main gloss component.

REFLECTED IMAGE QUALITY (RIQ)
RIQ is used to quantify effects such as orange peel and surface waviness. This new parameter gives higher resolution results compared to Distinctness of Image (DOI) measurement and better mimics human perception of surface texture, especially on high quality finishes such as automotive. Symptoms of poor RIQ: orange peel, brush marks, waviness or other structures visible on the surface. Reflected images are distorted.
Causes – Application problems, incorrect coating flow, coating viscosity too high/low, sag or flow of coating before curing, incorrect particle size/distribution, overspray, improper flash/recoat time, inter coat compatibility, incorrect cure times and cure temperature.

DISTINCTNESS OF IMAGE (DOI)
A measure of how clearly a reflected image will appear in a reflective surface.

PEAK SPECULAR REFLECTANCE (RSPEC)
RSpec is the peak reflectance measured over a very narrow angle. RSpec values are lower in surfaces with textures including orange peel and waviness. It can be used to identify subtle differences in smooth surfaces. RSpec is more sensitive to longer wavelength texture than DOI.

quick-report-app-logo-all-white

Quick Report App Connection

The Rhopoint Quick Report app instantly transforms your measurement data into clear, professional, branded reports directly from your instrument. It saves time, reduces errors and lets you share results with confidence in just a few taps.
  1. Open the Rhopoint Quick Report App in Google Chrome or on a compatible mobile device
  2. On the instrument, ensure Bluetooth is switched on and set the BT mode to DATA
  3. In the app, press the pairing/link button in the top right-hand corner
  4. Select Connect from the drop-down menu
  5. Select the instrument from the available devices
  6. When the connection button turns green, the instrument is connected and the app is ready to receive measurements
  7. Take measurements on the instrument *
  8. Received readings will be displayed in the app for review and reporting

* For full instructions on installation, device connection, measurement transfer, data export and report creation, refer to the Rhopoint Quick Report App Manual.

The Quick Report App is available for Windows, Mac, Linux, Android and iOS; it can be run in Google Chrome or installed on a compatible device. On first connection, users are prompted to register and activate the included app subscription.