Settings Menu

image description

image description

image description
Used to enter calibration and factory options

image description
Used to adjust the instrument time and date

image description
Used to access links to an online manual.

image description
Maintenance menu

UNITS

Change the instruments displayed temperature between Degrees Celsius or Degrees Fahrenheit.

COOLANT

Change the water coolant type between chiller mode and mains mode.
Chiller mode should be used If the instrument is being used with a recirculating water chiller. When a coolant is not required an internal bypass valve will open so that the water flow is not stopped. This avoids the pump in the chiller from being deadheaded that can lead to permanent damage.
Mains mode should be used when the instrument is being used with mains water. The internal bypass valve is deactivated so that only water is taken when required.