Aesthetix

Rhopoint Aesthetix is a modular dual-camera–based sensor that captures detailed images of a surface under controlled lighting and derives a wide range of appearance metrics from them: gloss, haze, sparkle, scratches, texture, waviness, defects and more. It connects to the host PC over USB and is exposed to Elements Hub through the device class aesthetix.

For the network-attached inline variant see Aesthetix Inline. Both variants share most of this guide, but the inline variant supports a reduced metric set and a single calibration type — the per-feature notes below call this out where it matters.

Workflow Summary

  1. Initialize lifecycle services
  2. Start a scan with deviceClass: aesthetix
  3. Connect to the discovered device
  4. Calibrate (white → black → height) before the first measurement of the session
  5. Trigger measurements with the metric groups your workflow needs
  6. Disconnect and shut down

Connect Parameters

Aesthetix devices accept no operation parameters on POST /v1/devices. Only the deviceIdentifier is required.

Calibration

The Aesthetix exposes three calibration types via the calibration endpoints:

calibrationKey Purpose Parameters
white Reflectance + visual contrast reference, using the white standard reflectance (0–1), visualContrast (0–1)
black Black specular and aspecular reference, using the black standard gloss (GU)
height Height reference, using the texture standard height (µm)

The Aesthetix Inline variant only exposes black. See Calibration for the recommended order, the physical setup for each target, and troubleshooting tips.

Available Measurements

The Aesthetix exposes one measurement trigger, measure, executed via `POST /v1/devices/{deviceId}/measurement-triggers`. Which metrics it actually computes is selected through the metricGroupKeys parameter — pass an array of the keys below:

Metric group key What it produces Notes
gloss Gloss value, visual gloss, gloss ROI, 3D gloss plot Requires valid black calibration
fastGloss Faster, lower-fidelity gloss value
haze Haze C, contrast haze, log-haze, MC-DOI, visual haze indoors/outside, haze ROIs
visualContrast Visual contrast value Requires valid white calibration
visualGloss Visual gloss value Requires valid white/black calibration
visualHaze Visual haze value
contrastHaze Contrast haze value
sparkle Sparkle density, graininess, RGB, visibility / area / brightness / hue arrays, sparkle albedo image, 7 LED sparkle images, 7 LED sparkle overlays
scratchLinear Scratch length, visibility, count, area, area-mean and length-mean for linear scratches; scratch image; horizontal and vertical overlays
scratchRadial Same metrics as scratchLinear but for radial scratches
cell Cell number, amplitude, size, std-dev, max/min, hill size, fill factor, texture R/Rc/Rv/Rh, roughness, watershed metrics, cell overlay, 3D texture plot
roughness Surface roughness
grit Grit metric
waviness Waviness, PCI waviness, tension, waviness image
pciWaviness PCI waviness value (subset of waviness)
tension Tension value (subset of waviness)
bloom Bloom metric
crossCut Cross-cut class (ASTM, ISO), percent, image, full overlay, found overlay Requires the cross-cut module
sharpness Sharpness metric
spot Spot image
shadeImages Six LED shade images No metric values — image stack only
surface Surface RGB mean, surface image (cropped)

The metricGroupKeys parameter accepts a list of these keys; pass only the groups you actually need. Each additional group adds processing time on the device side.

Measurement Parameters

In addition to metricGroupKeys, the measure trigger accepts:

Parameter Type Description
metricGroupKeys EnumArray Selects which metric groups participate in the measurement. See table above.
includeAttachments Boolean Whether to include attachments such as images in the result. Defaults to true.
imageFormat String Image encoding inside the container: image/png (default), image/jpeg, or image/x-raw. See Measurement Image Formats.

Available Commands

Beyond measurements and calibrations, the Aesthetix exposes one command via the commands endpoint:

Command identifier Purpose Parameters
updateCalibrationValues Update the stored calibration target values without performing a full calibration. Useful when only the target values for a known good calibration have changed. glossCalibrationTarget, visualContrastCalibrationTarget, … (see GET /commands)

Image Output

Device images embedded in a measurement container — surface, sparkle, scratch, cross-cut, cell, waviness, spot, shade — are PNG-encoded by default. Use the imageFormat parameter to request JPEG (smaller) or raw bytes (uncropped originals straight from the device kernel). See Measurement Image Formats for the binary layout of image/x-raw and parsing examples in C++.

Mock Variant

For development and CI without physical hardware, the hub offers the device class aesthetix-mock. It exposes the same endpoints and parameter shapes but does not return real images or metric values. See Mock Devices.