Calibrations

Calibration brings a connected device's measurement chain back into agreement with a known reference (white, black, or height target). For instruments such as Aesthetix, a valid calibration is a precondition for trustworthy measurements — running a measurement on an uncalibrated device returns numbers, but they are not metrologically meaningful. The hub tracks the last calibration time per device and exposes a status endpoint so clients can prompt for re-calibration before it goes stale.

Endpoints Overview

Method Endpoint Description
GET /v1/devices/{deviceId}/calibrations List calibrations available on the connected device
POST /v1/devices/{deviceId}/calibrations Execute a calibration
GET /v1/devices/{deviceId}/calibrations/status Get current calibration status per calibration type

Listing Available Calibrations

GET /v1/devices/{deviceId}/calibrations
curl http://localhost:42042/v1/devices/12f1d7dd07ac42a088c8f961b39d68ff/calibrations

Returns the device's calibration definitions, including the parameters each one accepts. Example shape:

[
  {
    "identifier": "white",
    "title": "White Calibration",
    "description": "Performs white reference calibration using the white calibration standard.",
    "category": "calibration",
    "parameters": [
      { "key": "reflectance",    "type": "Double", "description": "Percent, 0.42 ≙ 42 %" },
      { "key": "visualContrast", "type": "Double", "description": "Percent, 0.42 ≙ 42 %" }
    ]
  }
]

The set of calibrations is device-class–specific. For example, an Aesthetix supports white, black and height, an Aesthetix Inline only black. See Aesthetix-specific calibration details.

Executing a Calibration

POST /v1/devices/{deviceId}/calibrations
Content-Type: application/json

{
  "calibrationKey": "white",
  "parameters": [
    { "key": "reflectance",    "value": 0.95 },
    { "key": "visualContrast", "value": 0.92 }
  ]
}
curl -X POST http://localhost:42042/v1/devices/12f1d7dd07ac42a088c8f961b39d68ff/calibrations \
  -H "Content-Type: application/json" \
  -d '{
    "calibrationKey": "white",
    "parameters": [
      { "key": "reflectance",    "value": 0.95 },
      { "key": "visualContrast", "value": 0.92 }
    ]
  }'

Request Body

Field Type Required Description
calibrationKey string Yes Identifier from GET /v1/devices/{deviceId}/calibrations (e.g. white, black, height).
parameters array No Calibration parameters as { "key": "...", "value": ... } entries. Keys and value types are device-class–specific (see GET response).

Response

{
  "calibrationKey": "white",
  "success": true
}

The call is synchronous and only returns once the calibration has completed on the device. Calibration can take several seconds; clients should set a generous HTTP timeout (≥30 s).

The physical calibration target must be in place before the request is sent. Calibrating against the wrong target will produce a successful response but invalid downstream measurements. See the device-specific calibration page for which target to use with which calibrationKey.

Getting the Calibration Status

GET /v1/devices/{deviceId}/calibrations/status

Returns the current state of every calibration type the device supports:

{
  "deviceId": "12f1d7dd07ac42a088c8f961b39d68ff",
  "deviceSerialNumber": "AXI8000027",
  "calibrations": [
    {
      "calibrationKey": "white",
      "lastCalibrationTime": "2025-11-10T07:32:14.000Z",
      "status": "CalibrationValid",
      "recommendedInterval": "2.00:00:00",
      "overdueInterval": "7.00:00:00"
    },
    {
      "calibrationKey": "black",
      "lastCalibrationTime": null,
      "status": "NotCalibrated",
      "recommendedInterval": "2.00:00:00",
      "overdueInterval": "7.00:00:00"
    }
  ]
}

Status Values

Status Meaning Recommended action
NotCalibrated No calibration of this type has ever been performed for this device's serial number. Calibrate before measuring.
CalibrationValid Last calibration is younger than recommendedInterval. Continue measuring.
CalibrationRecommended Last calibration is older than recommendedInterval but younger than overdueInterval. Calibrate at the next convenient point.
CalibrationOverdue Last calibration is older than overdueInterval. Calibrate before further measurements.

The intervals follow .NET TimeSpan formatting (d.hh:mm:ss). Typical defaults are 2 days (recommended) and 7 days (overdue), but the device definition is authoritative — read the values from the response rather than hard-coding them.

Persistence

Calibration data is stored per serial number + device class, not per session identifier. Disconnecting and re-connecting a device — or restarting the hub — does not invalidate previously persisted calibrations. The storage location is the dataDirectory passed to `POST /v1/lifecycle/initialize`.

Best Practices

  • After connecting a device, call GET /v1/devices/{deviceId}/calibrations/status and surface any NotCalibrated, CalibrationRecommended or CalibrationOverdue entries to the operator before allowing measurements to start.
  • Calibrate in the order recommended by the device manufacturer (typically Black → White → Height for Aesthetix). Some calibrations depend on earlier ones being valid.
  • Read parameters from GET /v1/devices/{deviceId}/calibrations and only send the values listed there — sending unknown keys is silently ignored, which makes typos hard to spot.
  • Time the calibration prompts using the status endpoint, not your own scheduler — the intervals can change between firmware updates.

Error Cases

Condition HTTP Error Code
Connected device not found 404 E5
Unknown calibrationKey for this device 404 E28
Calibration timed out (hardware likely disconnected) 502 E29

See Error Handling for the response format.