# Main Operation > [!note] 此页面尚未翻译,以英文显示。 Main screen overview ![image description](../_images/1756291072497-Main-Streen-with-labels.webp) The main operating screen for the MFFT instrument will look like the above. In this example range 3 has been selected but the instrument is not set to start achieving target temperature. To change the target range, press the back or return button ![image description](../_images/1756291111710-back.webp) ![image description](../_images/1756291187457-range-select.webp) Press the left of right arrow buttons to cycle through the range options. ![image description](../_images/1756291159526-arrow-L.webp) ![image description](../_images/1756291163958-arrow-r.webp) The MFFT instrument has 10 operating ranges. Range 0 – 9. Target Cold end and Hot end for each range is shown in the table below. ![image description](../_images/1756291299155-Temperature-ranges.webp) When the desired range has been selected use the tick button to confirm the selection. ![image description](../_images/1756291348921-tick.webp) ![image description](../_images/1756291388472-operating-screen.webp) With the desired range has been selected press the play button. ![image description](../_images/1756291429299-play.webp) The heating and cooling process will start. The play button will change to a pause button. ![image description](../_images/1756291435845-pause.webp) Pressing the pause button will stop the heating and cooling process. When the target temperature is reached the thermometer outline will turn green, indicating correct temperature. ![image description](../_images/1756291727315-operating-screen-ready.webp) An audio system ready beep will also sound. To mute the audio alert the speaker icon can be pressed. ![image description](../_images/1756291752647-sound.webp) Pressing the right-hand double arrow button will progress to the next screen. ![image description](../_images/1756291792403-forward.webp) The plate temperature will continue to maintain at the set points. The sample to be tested can now be placed on the test pattern. > [!tip] For sample application, see [Applying the Sample](rhopoint-mfft-applying-the-sample.md). ![image description](../_images/1756291981363-result-screen.webp) The screen will now show the hot and cold set points and the MFFT point.\ When the sample has dried, move the cursor to the MFFT point. ![image description](../_images/1756292030768-test-bed-with-sample.jpg) When the cursor is at the correct position press the tick button to accept the values. ![image description](../_images/1756291348921-tick.webp) > [!note] > The tick button will only be available if the MFFT instrument is at the correct temperatures for the range selected. ![image description](../_images/1756292118693-results.webp) The results of the test are displayed on the screen. Press the return button to exit back to the main range stabilization screen. ![image description](../_images/1756291111710-back.webp) Press the save button to export the results to a CSV file. ![image description](../_images/1756292154168-save.webp) A csv file will be saved on a USB drive if inserted. The USB port is on the rear of the instrument. **Example of the CSV file** Serial Number,MFFT123456\ Date,21 Aug 2025\ Time,13:52\ Range,9\ Cold,63\ Hot,90\ MFFT Point,72 > [!tip] The file is a comma-separated CSV.