# Calibrations > [!note] Esta página ainda não foi traduzida e é apresentada em inglês. Calibration brings a connected device's measurement chain back into agreement with a known reference (white, black, or height target). For instruments such as Aesthetix, a valid calibration is a precondition for trustworthy measurements — running a measurement on an uncalibrated device returns numbers, but they are not metrologically meaningful. The hub tracks the last calibration time per device and exposes a `status` endpoint so clients can prompt for re-calibration before it goes stale. ### Endpoints Overview | Method | Endpoint | Description | | --- | --- | --- | | GET | `/v1/devices/{deviceId}/calibrations` | List calibrations available on the connected device | | POST | `/v1/devices/{deviceId}/calibrations` | Execute a calibration | | GET | `/v1/devices/{deviceId}/calibrations/status` | Get current calibration status per calibration type | ### Listing Available Calibrations ```http GET /v1/devices/{deviceId}/calibrations ``` ```bash curl http://localhost:42042/v1/devices/12f1d7dd07ac42a088c8f961b39d68ff/calibrations ``` Returns the device's calibration definitions, including the parameters each one accepts. Example shape: ```json [ { "identifier": "white", "title": "White Calibration", "description": "Performs white reference calibration using the white calibration standard.", "category": "calibration", "parameters": [ { "key": "reflectance", "type": "Double", "description": "Percent, 0.42 ≙ 42 %" }, { "key": "visualContrast", "type": "Double", "description": "Percent, 0.42 ≙ 42 %" } ] } ] ``` The set of calibrations is device-class–specific. For example, an Aesthetix supports `white`, `black` and `height`, an Aesthetix Inline only `black`. See [Aesthetix-specific calibration details](rhopoint-elements-hub-instruments-aesthetix-calibration.md). ### Executing a Calibration ```http POST /v1/devices/{deviceId}/calibrations Content-Type: application/json { "calibrationKey": "white", "parameters": [ { "key": "reflectance", "value": 0.95 }, { "key": "visualContrast", "value": 0.92 } ] } ``` ```bash curl -X POST http://localhost:42042/v1/devices/12f1d7dd07ac42a088c8f961b39d68ff/calibrations \ -H "Content-Type: application/json" \ -d '{ "calibrationKey": "white", "parameters": [ { "key": "reflectance", "value": 0.95 }, { "key": "visualContrast", "value": 0.92 } ] }' ``` #### Request Body | Field | Type | Required | Description | | --- | --- | --- | --- | | `calibrationKey` | string | Yes | Identifier from `GET /v1/devices/{deviceId}/calibrations` (e.g. `white`, `black`, `height`). | | `parameters` | array | No | Calibration parameters as `{ "key": "...", "value": ... }` entries. Keys and value types are device-class–specific (see `GET` response). | #### Response ```json { "calibrationKey": "white", "success": true } ``` The call is synchronous and only returns once the calibration has completed on the device. Calibration can take several seconds; clients should set a generous HTTP timeout (≥30 s). > [!warning] > The physical calibration target must be in place **before** the request is sent. Calibrating against the wrong target will produce a successful response but invalid downstream measurements. See the device-specific calibration page for which target to use with which `calibrationKey`. ### Getting the Calibration Status ```http GET /v1/devices/{deviceId}/calibrations/status ``` Returns the current state of every calibration type the device supports: ```json { "deviceId": "12f1d7dd07ac42a088c8f961b39d68ff", "deviceSerialNumber": "AXI8000027", "calibrations": [ { "calibrationKey": "white", "lastCalibrationTime": "2025-11-10T07:32:14.000Z", "status": "CalibrationValid", "recommendedInterval": "2.00:00:00", "overdueInterval": "7.00:00:00" }, { "calibrationKey": "black", "lastCalibrationTime": null, "status": "NotCalibrated", "recommendedInterval": "2.00:00:00", "overdueInterval": "7.00:00:00" } ] } ``` #### Status Values | Status | Meaning | Recommended action | | --- | --- | --- | | `NotCalibrated` | No calibration of this type has ever been performed for this device's serial number. | Calibrate before measuring. | | `CalibrationValid` | Last calibration is younger than `recommendedInterval`. | Continue measuring. | | `CalibrationRecommended` | Last calibration is older than `recommendedInterval` but younger than `overdueInterval`. | Calibrate at the next convenient point. | | `CalibrationOverdue` | Last calibration is older than `overdueInterval`. | Calibrate before further measurements. | The intervals follow .NET `TimeSpan` formatting (`d.hh:mm:ss`). Typical defaults are 2 days (recommended) and 7 days (overdue), but the device definition is authoritative — read the values from the response rather than hard-coding them. ### Persistence Calibration data is stored per **serial number + device class**, not per session identifier. Disconnecting and re-connecting a device — or restarting the hub — does not invalidate previously persisted calibrations. The storage location is the `dataDirectory` passed to [`POST /v1/lifecycle/initialize`](rhopoint-elements-hub-endpoints-functionality-lifecycle.md). ### Best Practices - After connecting a device, call `GET /v1/devices/{deviceId}/calibrations/status` and surface any `NotCalibrated`, `CalibrationRecommended` or `CalibrationOverdue` entries to the operator before allowing measurements to start. - Calibrate in the order recommended by the device manufacturer (typically Black → White → Height for Aesthetix). Some calibrations depend on earlier ones being valid. - Read `parameters` from `GET /v1/devices/{deviceId}/calibrations` and only send the values listed there — sending unknown keys is silently ignored, which makes typos hard to spot. - Time the calibration prompts using the `status` endpoint, not your own scheduler — the intervals can change between firmware updates. ### Error Cases | Condition | HTTP | Error Code | | --- | --- | --- | | Connected device not found | 404 | `E5` | | Unknown `calibrationKey` for this device | 404 | `E28` | | Calibration timed out (hardware likely disconnected) | 502 | `E29` | See [Error Handling](rhopoint-elements-hub-error-handling.md) for the response format.