# Calibrating Aesthetix in AE > [!note] Deze pagina is nog niet in het Nederlands beschikbaar en wordt in het Engels weergegeven. Once a module is chosen, it may be required to calibrate the sensor. ## Calibration Standards The following standards must be used dependent on the module. | Module | Standard | |------------------------|-----------------------------------| | Gloss Module | B8000-011 Gloss Module Standard | | Texture Module | B8000-012 Texture Module Standard | | Coatings Physical Test | B8000-011 Gloss Module Standard | Modules which do not require calibration include: - Sparkle Module Click on the calibration icon to begin calibration and follow the on-screen instructions. [Calibrating an Instrument in AE](rhopoint-appearance-elements-calibrating-an-instrument-in-ae.md) > [!tip] Calibration interval > The instrument should be checked by measuring the calibration tile daily and comparing read values with the certified values. If values are out of tolerance, recalibrate the sensor. > [!warning] Calibration artifacts must be clean with no contamination, visible damage, or fingerprints.