# Calibrate the TAMS > [!note] このページはまだ翻訳されておらず、英語で表示されています。 ## Calibrate the TAMS Before taking measurements, calibrate TAMS using the supplied plate so focus and reference values are correctly calibrated. ### Calibration plate The plate includes three tiles: - **Plastic‑ref** – for surface focus. - **Silver‑ref** – for screen focus and reference calibration. - **Check tile‑ref** – for later verification only (not used during calibration). ### Start calibration 1. On the instrument, select **Menu → Calibration → Start calibration process**. 2. Confirm with **YES** and press **OK**. ### Step 1 – Plastic‑ref 1. Place TAMS on the **Plastic‑ref** tile (top tile). 2. Ensure all four feet are flat on the tile. 3. Select **Continue** and press **OK**.ly calibrated TAMS now sets the surface autofocus. ### Step 2 – Silver‑ref 1. Move TAMS to the **Silver‑ref** tile (mirror‑like middle tile). 2. Ensure all four feet are flat on the tile. 3. Select **Continue** and press **OK**. TAMS now sets the screen autofocus and reference calibration, then returns to normal operation. ### Step 2 – Gloss 1. Move TAMS to the **Gloss‑ref** tile. 2. Ensure all four feet are flat on the tile. 3. Select **Continue** and press **OK**. TAMS now calibrates the gloss measurement.