# Calibration > [!note] このページはまだ翻訳されておらず、英語で表示されています。 The Aesthetix is calibrated via the generic [calibration endpoints](rhopoint-elements-hub-endpoints-functionality-calibrations.md) — `GET/POST /v1/devices/{deviceId}/calibrations` and `GET /calibrations/status`. This page covers the Aesthetix-specific calibration types, what physical target each one needs, and the recommended order. ### Available Calibration Types | `calibrationKey` | Title | Physical target | Parameters | | --- | --- | --- | --- | | `white` | White Calibration | White calibration standard | `reflectance` (0–1), `visualContrast` (0–1) | | `black` | Black Calibration | Black calibration standard | `gloss` (GU) | | `height` | Height Calibration | Texture standard | `height` (µm) | Every calibration is valid for **two days** (recommended interval) and goes overdue after **seven days**. The hub tracks the timestamps per device serial number — see [`GET /calibrations/status`](rhopoint-elements-hub-endpoints-functionality-calibrations.md#getting-the-calibration-status). > [!note] > The Aesthetix Inline variant only exposes `black`. The `white` and `height` calibrations are not available there. ### Recommended Order Run the calibrations in this sequence the first time a device is calibrated, after a long idle period, or whenever the status endpoint reports `NotCalibrated` / `CalibrationOverdue`: 1. **Black** — establishes the optical zero for the specular and aspecular paths. 2. **White** — references the reflectance and visual-contrast scales against the white standard. 3. **Height** — references the height scale against the texture standard. White and height depend on a valid black calibration; running them first against an outdated black reference yields wrong scaling. The hub does not enforce this order, so the integrator's UI or workflow must. ### White Calibration Used to reference reflectance and visual-contrast measurements against the white calibration standard supplied with the instrument. ```http POST /v1/devices/{deviceId}/calibrations Content-Type: application/json { "calibrationKey": "white", "parameters": [ { "key": "reflectance", "value": 0.95 }, { "key": "visualContrast", "value": 0.92 } ] } ``` | Parameter | Unit | Notes | | --- | --- | --- | | `reflectance` | Fraction (`0.95` ≙ 95 %) | The reflectance value printed on the white standard. | | `visualContrast` | Fraction (`0.92` ≙ 92 %) | The visual-contrast value printed on the white standard. | **Before sending the request:** place the white standard over the measurement window with the white surface facing the optics. Hold the instrument steady — the device captures multiple frames during calibration. ### Black Calibration Establishes the optical zero. The Aesthetix performs both **specular** and **aspecular** black calibration in a single request; the Aesthetix Inline only performs specular. ```http POST /v1/devices/{deviceId}/calibrations Content-Type: application/json { "calibrationKey": "black", "parameters": [ { "key": "gloss", "value": 0.5 } ] } ``` | Parameter | Unit | Notes | | --- | --- | --- | | `gloss` | GU (Gloss Units) | The gloss value printed on the black standard. | **Before sending the request:** place the black standard over the measurement window with the black surface facing the optics. ### Height Calibration References the height (texture) scale against the texture standard. ```http POST /v1/devices/{deviceId}/calibrations Content-Type: application/json { "calibrationKey": "height", "parameters": [ { "key": "height", "value": 50.0 } ] } ``` | Parameter | Unit | Notes | | --- | --- | --- | | `height` | µm | The reference height of the texture standard, in micrometres. | **Before sending the request:** place the texture standard over the measurement window. Ensure it sits flat and centered. ### Verifying the Result After the call returns successfully, query the status to confirm the new timestamp: ```http GET /v1/devices/{deviceId}/calibrations/status ``` A freshly executed calibration moves to `CalibrationValid` and its `lastCalibrationTime` matches the request time (within network skew). If the status remains `NotCalibrated` after a successful response, the calibration parameter values were likely ignored — re-check the `key` names against `GET /v1/devices/{deviceId}/calibrations`. ### Troubleshooting | Symptom | Likely cause | Fix | | --- | --- | --- | | `E29 Calibration timeout` | The instrument disconnected during the calibration sequence. | Reconnect the device (re-scan + connect) and retry. | | Calibration succeeds but downstream measurements are clearly wrong | Wrong physical target was on the window, or the target values printed on the standard were entered with the wrong scale (e.g. `95` instead of `0.95`). | Re-calibrate with the correct target and verify `reflectance`/`visualContrast` are passed as fractions, not percentages. | | Status stays `NotCalibrated` after a 200 OK response | Parameter `key` was misspelled and ignored. | Read the accepted keys from `GET /v1/devices/{deviceId}/calibrations` and retry. | | Calibration takes longer than the HTTP client timeout | Default client timeouts (often 5–10 s) are too short for calibration. | Raise the per-request timeout to at least 30 s. | ### Error Cases | Condition | HTTP | Error Code | | --- | --- | --- | | Connected device not found | 404 | `E5` | | Unknown `calibrationKey` | 404 | `E28` | | Calibration timed out (likely hardware disconnect) | 502 | `E29` | See [Error Handling](rhopoint-elements-hub-error-handling.md) for the full response format.